Microfluidics · 2019
Modular microfluidic systems cast from 3D-printed molds for imaging leukocyte adherence to differentially treated endothelial cultures
Microfluidic systems are very useful for in vitro studies of interactions between blood cells and vascular endothelial cells under flow, and several commercial solutions exist. However, the availability of customizable, user-designed devices is largely restricted to researchers with expertise in photolithography and access to clean room facilities.
Performance the paper reports
0 of 5- Motion—Motion accuracy: not reported
- Volume—Volumetric accuracy: not reported
- Viability—Cell viability: not reported
- Unattended—Unattended operation: not reported
- Standard—Standard compliance: not reported
The full text was read and states no figure for any of the five.
How hard this is to adopt
2.4 of 5, higher is easierEach score is shown against the range the other 59 builds actually occupy, because no criterion uses the full 1–5 scale. These describe how reproducible the build is, not how good the science is.
Resolution
2/1–5Smallest feature/positioning evidence about 250 µm.
Throughput
2/2–5Operation appears single-unit or manually constrained.
Parts sourcing
2/2–4Requires specialized or custom-built components.
Skill needed
2/2–4Multiple advanced technical skills appear required.
Facilities needed
2/2–4Requires specialized facilities, equipment, or consumables.
Documentation
3/2–5Some open documentation or files are available.
Validation effort
4/2–4Validation includes standards, benchmarking, replicates, or multi-condition tests.
Cycle time
not scoredNo clear speed comparison found
2 further criteria are recorded but not shown
- Application Level — scored 4 (75% of records share one value), so it cannot separate one build from another.
- Build Time (scored for only 6 of 60 records), so it cannot separate one build from another.
Both remain in the published dataset.
Have you built this?
Report what it took, including attempts that stalled. A failed build tells the next laboratory what the documentation leaves out.
Know what this record is missing?
The index has no cost, licence on the design files or link to the design files for this build. If you are an author or have built it, you know; a GitHub issue pre-filled with this record takes a minute.
Record metadata
Publication data
Matched by doi · retrieved 2026-07-21
Abstract
Microfluidic systems are very useful for in vitro studies of interactions between blood cells and vascular endothelial cells under flow, and several commercial solutions exist. However, the availability of customizable, user-designed devices is largely restricted to researchers with expertise in photolithography and access to clean room facilities. Here we describe a strategy for producing tailor-made modular microfluidic systems, cast in PDMS from 3D-printed molds, to facilitate studies of leukocyte adherence to endothelial cells. A dual-chamber barrier module was optimized for culturing two endothelial cell populations, separated by a 250 μm wide dividing wall, on a glass slide. In proof-of-principle experiments one endothelial population was activated by TNFα, while the other served as an internal control. The barrier module was thereafter replaced with a microfluidic flow module, enclosing both endothelial populations in a common channel. A suspension of fluorescently-labeled leukocytes was then perfused through the flow module and leukocyte interactions with control and TNFα-treated endothelial populations were monitored in the same field of view. Time-lapse microscopy analysis confirmed the preferential attachment of leukocytes to the TNFα-activated endothelial cells. We conclude that the functionality of these modular microfluidic systems makes it possible to seed and differentially activate adherent cell types, and conduct controlled side-by-side analysis of their capacity to interact with cells in suspension under flow. Furthermore, we outline a number of practical considerations and solutions associated with connecting and switching between the microfluidic modules, and the advantages of simultaneously and symmetrically analyzing control and experimental conditions in such a microfluidic system.
Bibliographic details
Access and metrics
Topics and keywords
Perspective and practical signals
Why it matters
No narrative note curated yet.
Limitations
No limitation note curated yet.