Liquid Handling · Pipetting/dispensing · Commit: Digital Pipette · 2026 · Digital Discovery
Commit: Digital pipette: open hardware for liquid transfer in self-driving laboratories
Robot-mounted 3D-printed digital pipette for liquid transfer in self-driving labs
Performance the paper reports
2 of 5- Motion—Motion accuracy: not reported
- Volume0.2%Volumetric accuracy: reported
- Viability—Cell viability: not reported
- Unattended—Unattended operation: not reported
- StandardISO 8655Standard compliance: reported
- Volumetric accuracy0.2%
[curated metric] 0.2% random error at 10 mL nominal volume (ISO 8655-6); meets ISO 8655-2 accuracy range
- Reported against a written standardISO 8655
[curated metric] 0.2% random error at 10 mL nominal volume (ISO 8655-6); meets ISO 8655-2 accuracy range
How hard this is to adopt
2.9 of 5, higher is easierEach score is shown against the range the other 59 builds actually occupy, because no criterion uses the full 1–5 scale. These describe how reproducible the build is, not how good the science is.
Resolution
1/1–5Nominal 10 mL air-displacement syringe; ISO tests at 1/5/10 mL with smallest tested 0.2 mL; authors concede uL-level accuracy needs higher-precision actuators.
Throughput
2/2–5Single-channel pipette handling one liquid at a time; automated tip exchange but external tip remover and greasing upkeep; no plate or full workflow demonstrated.
Parts sourcing
3/2–4Mix of common parts and custom/printed components.
Skill needed
4/2–4Build/operation described as low-skill or basic assembly.
Facilities needed
2/2–4Requires specialized facilities, equipment, or consumables.
Documentation
4/2–5Open resources include several build or operation artifacts.
Validation effort
4/2–4ISO 8655-6 gravimetric at three volumes against v1 and ISO limits, plus explicit benchmark against four human pipettors (Levene, p=0.0046); calibration curves given.
Cycle time
3/1–4No timing comparison reported; only a 10-minute assembly time, with no aspiration/dispense or cycle-time data versus manual or commercial pipetting.
2 further criteria are recorded but not shown
- Application Level — scored 4 (75% of records share one value), so it cannot separate one build from another.
- Build Time — scored 5 (scored for only 6 of 60 records), so it cannot separate one build from another.
Both remain in the published dataset.
Have you built this?
Report what it took, including attempts that stalled. A failed build tells the next laboratory what the documentation leaves out.
Know what this record is missing?
The index has no licence on the design files for this build. If you are an author or have built it, you know; a GitHub issue pre-filled with this record takes a minute.
Record metadata
Publication data
Matched by doi · retrieved 2026-07-21
Abstract
We present an updated version of the digital pipette, an economical 3D-printed device for accurate liquid transfers by robotic arms. The new version supports disposable tip replacement to reduce contamination while maintaining high accuracy.
Bibliographic details
Access and metrics
Topics and keywords
Funding
- Canada First Research Excellence Fund · CFREF-2022-00042
- Japan Society for the Promotion of Science · JP25K21333
Perspective and practical signals
Why it matters
Open hardware for accurate liquid transfer in self-driving labs; reports accuracy within ISO 8655-2 permissible error.
Limitations
Requires a robot arm/gripper to operate -- not a standalone bench tool